m8ta
You are not authenticated, login.
text: sort by
tags: modified
type: chronology
{549} is owned by tlh24.
{1018}
hide / / print
ref: Rouse-2011.06 tags: BMI chronic DBS bidirectional stimulator Washington Medtronic ASIC translational date: 03-05-2012 23:56 gmt revision:3 [2] [1] [0] [head]

PMID-21543839[0] A chronic generalized bi-directional brain-machine interface.

  • Using a commercial neurostimulator package & battery etc.
  • "A key goal of this research prototype is to help broaden the clinical scope and acceptance of NI techniques, particularly real-time brain state detection" Good purpose! good work!
  • Augments the stimulator with 4 channels of ECoG/LFP + accelerometer + wireless telemetry.
    • Can be used to detect parkinsons state or pre-epileptiform behavior.
      • Much of this has been though of before, it just took the technology to catch up & a group to make it.
    • Chronic data is needed from humans -- animal models are often inadequate.
  • Tested in a primate for brain control of a cursor: 1D control using ECoG.
    • Good Left/right ROC, actually.
    • A large cost is simply the clinical testing; hence they piggy-back on an existing design.
    • There should be more research-industry collaborations like this.
  • impressive specs.
  • SVM classification algorithm (only consumed 10uW!) for data compression.
  • short-time Fourier transform for extracting the power over a given band. This using a modified chopper-amplification scheme. Output data has a bandwidth of less than 5Hz, which greatly reduces processing requirements.
  • Lots of processing on the BASIC chip, much like here.
  • Also see the press release

____References____

[0] Rouse AG, Stanslaski SR, Cong P, Jensen RM, Afshar P, Ullestad D, Gupta R, Molnar GF, Moran DW, Denison TJ, A chronic generalized bi-directional brain-machine interface.J Neural Eng 8:3, 036018 (2011 Jun)

{831}
hide / / print
ref: Blum-2004.01 tags: microstimulator MEA Georgia Blum integrated circuit date: 01-03-2012 06:53 gmt revision:3 [2] [1] [0] [head]

PMID-17271195[0] Models of stimulation artifacts applied to integrated circuit design.

  • for MEAs.
  • Idea: develop a model of the stimulation artifact so they can optimize removal in SPICE.
  • reference documents that say that biphasic stimulation + active artifact suppression (by discharging the electrodes after stimulation, [1]) are acknowledged means of reducing stimulus artifact.
  • artifact appears to be 1ms saturating, 6ms non-saturating pulse.
  • a little light on details.

____References____

[0] Blum R, Ross J, Das S, Brown E, Deweerth S, Models of stimulation artifacts applied to integrated circuit design.Conf Proc IEEE Eng Med Biol Soc 6no Issue 4075-8 (2004)
[1] Jimbo Y, Kasai N, Torimitsu K, Tateno T, Robinson HP, A system for MEA-based multisite stimulation.IEEE Trans Biomed Eng 50:2, 241-8 (2003 Feb)

{944}
hide / / print
ref: Hassell-2007.08 tags: microstimulator ICMS ASIC Hassel date: 12-20-2011 06:17 gmt revision:1 [0] [head]

IEEE-4352820 (pdf) Constant-Current Adjustable-Waveform Microstimulator for an Implantable Hybrid Neural Prosthesis

  • Focus on hybrid electro-chemical prosthesis: electrical release of glutamate. For "native output modality".
    • For use in neuronal culture, at least initially.
  • Talk about constant-current sources, typically current-mirrors. Want the following:
    • Linearity
    • High output impedance.
    • voltage compliance.
    • Are these things necessarily well motivated? constant power may be better. eh.
  • WAnt to cascode current output for greater output impedance.
  • They appear not to have completed the ASIC (at time of publication).

____References____

Hassell, T.J. and Jedlicka, S.S. and Rickus, J.L. and Irazoqui, P.P. Engineering in Medicine and Biology Society, 2007. EMBS 2007. 29th Annual International Conference of the IEEE 2436 -2439 (2007)

{226}
hide / / print
ref: work notes-0 tags: web stimulator SUNY ICMS python webinterface project date: 03-26-2007 04:26 gmt revision:1 [0] [head]

we are proud of this :)