ref: Brown-2008.03 tags: microstimulation recording artifact supression MEA ICMS date: 12-28-2011 20:43 gmt revision:3 [2] [1] [0] [head]

IEEE-4464125 (pdf) Stimulus-Artifact Elimination in a Multi-Electrode System

  • Stimulate and record on the same electrode within 3ms; record on adjacent electrodes within 500us.
  • Target at MEAs, again.
  • Notes that very small charge mismatches of 1% or less, which is common and acceptable in traditional analog circuit designs, generates an artifact that saturates the neural amp signal chain.
  • for stimulating & recording on the same electrode, the the residual charge must be brought down to 1/1e5 the stimulating charge (or less).
  • paper follows upon {833} -- shared author, Blum -- especially in the idea of using active feedback to cancel artifact charge & associated voltage.
  • target the active feedback for keeping all amplifier out of saturation.
  • vary highpass filter poles during artifact supression (!)
  • bias currents of 1fA on the feedback highpass stage. yikes.

Brown EA, Ross JD, Blum RA, Yoonkey N, Wheeler BC, and DeWeerth SP (2008) Stimulus-Artifact Elimination in a Multi-Electrode System. IEEE TRans. Biomed. Circuit Sys. 2. 10-21