{832} revision 2 modified: 12-17-2011 01:41 gmt

PMID-12665038[0] A system for MEA-based multisite stimulation.

  • stimulate and record the same MEA channel.
  • used voltage-control stimulation.
  • very low leakage current switches, DG202CSE, 100Gohm, Maxim, above. non-mechanical = low vibration.
  • switches switch between stimulator and preamp. obvious.
  • uses active shorting post-stimulation to remove residual charge,
  • uses active sample/hold of the preamplifier while the stimulator is connected to the electrodes.
  • adds stimulation pulse to the initial electrode offset (interesting!)

____References____

[0] Jimbo Y, Kasai N, Torimitsu K, Tateno T, Robinson HP, A system for MEA-based multisite stimulation.IEEE Trans Biomed Eng 50:2, 241-8 (2003 Feb)