{744} revision 2 modified: 01-03-2012 01:07 gmt

PMID-19042063[0] Technology and instrumentation for detection and conditioning of the surface electromyographic signal: state of the art

  • good background & review of surface EMG (sEMG) - noise levels, electrodes, electronics. eg. Instrumentation amplifiers with an input resistance < 100MOhm are not recommended, and the lower the input capacitance, the better: the impedance of a 10pf capacitor at 100hz is 160MOhm.
  • Low and balanced input impedances are required to reduce asymmetric filtering of common-mode power-line noise.

____References____

[0] Merletti R, Botter A, Troiano A, Merlo E, Minetto MA, Technology and instrumentation for detection and conditioning of the surface electromyographic signal: state of the art.Clin Biomech (Bristol, Avon) 24:2, 122-34 (2009 Feb)