Ultraflexible nanoelectronic probes form reliable, glial scar–free neural integration

  • SU-8 asymptotic H2O absorption is 3.3% in PBS -- quite a bit higher than I expected, and higher than PI.
  • Faced yield problems with contact litho at 2-3um trace/space.
  • Good recordings out to 4 months!
  • 3 minutes / probe insertion.
  • Fab:
    • Ni release layer, Su-8 2000.5. "excellent tensile strength" --
      • Tensile strength 60 MPa
      • Youngs modulus 2.0 GPa
      • Elongation at break 6.5%
      • Water absorption, per spec sheet, 0.65% (but not PBS)
    • 500nm dielectric; < 1% crosstalk; see figure S12.
    • Pt or Au rec sites, 10um x 20um or 30 x 30um.
    • FFC connector, with Si substrate remaining.
  • Used transgenic mice, YFP expressed in neurons.
  • CA glue used before metabond, followed by Kwik-sil silicone.
  • Neuron yield not so great -- they need to plate the electrodes down to acceptable impedance. (figure S5)
    • Measured impedance ~ 1M at 1khz.
  • Unclear if 50um x 1um is really that much worse than 10um x 1.5um.
  • Histology looks realyl great, (figure S10).
  • Manuscript did not mention (though the did at the poster) problems with electrode pull-out; they deal with it in the same way, application of ACSF.