ref: Salcman-1973.07 tags: Salcman MEA microelectrodes chronic recording glass cyanocrylate date: 12-29-2017 04:33 gmt revision:7 [6] [5] [4] [3] [2] [1] [head]

PMID-4708761 Design, Fabrication, and In Vivo Behavior of Chronic Recording Intracortical Microelectrodes

  • Teflon-coated 25um Pt-Ir (90/10)
  • Heat fuse this with a glass micropipette & backfill with cyanoacrylate. {1011}
    • Isobutyl acrylate is hydrolysed more slowly and hence is less toxic to the surronding tissue
    • cyanoacrylate is apparently biodegradable.
  • Durable, stable: one electrode displayed a single cortical spike (though not necessarily the same one) for more than 90 consecutive days.
  • unacceptably low impedance = 100K or less
  • Unit activity was present only 10-24H after surgery.
  • formal review of even older microelectrode studies.
  • 10nA should be 100x too small to have any effect on a platinum tip [17]
  • A seperable cell with a SNR of 3:1 would become lost if the electrode tip moved 15um away from a 20um soma.
    • "It becomes clear that the problem of holding single units for prolonged periods in the unrestrained animal is not achieved without considerable difficulty". Yet they think they have solved it.

____References____

Salcman, Michael and Bak, Martin J. Design, Fabrication, and In Vivo Behavior of Chronic Recording Intracortical Microelectrodes Biomedical Engineering, IEEE Transactions on BME-20 4 253 -260 (1973)