m8ta
you are not logged in, login. new entry
text: sort by
tags: modified
type: chronology
{1176}
hide / edit[5] / print
ref: Gilgunn-2012 tags: kozai neural recording electrodes compliant parylene flexible dissolve date: 02-24-2017 19:14 gmt revision:5 [4] [3] [2] [1] [0] [head]

IEEE-6170092 (pdf) An ultra-compliant, scalable neural probe with molded biodissolvable delivery vehicle

    • Optical coherence tomography is cool.
  • Large footprint - 150 or 300um, 135um thick (13500 or 40500 um^2; c.f. tungsten needle 1963 (50um) or 490 (25um) um^2.)
  • Delivery vehicle is fabricated from biodissolvable carboxy-methylcellulose (CMC).
    • Device dissolves within three minutes of implantation.
    • Yet stiff enough to penetrate the dura of rats (with what degree of dimpling?)
    • Lithographic patterning process pretty clever, actually.
    • Parylene-X is ~ 1.1 um thick.
    • 500nm Pt is patterned via ion milling with a photoresist mask.
    • Use thin 20nm Cr etch mask for both DRIE (STS ICP) and parylene etch.
  • Probes are tiny -- 10um wide, 2.7um thick, coated in parylene-X.
  • CMC polymer tends to bend and warp due to stress -- must be clamped in a special jig.
  • No histology. Follow-up?