m8ta
you are not logged in, login. new entry
text: sort by
tags: modified
type: chronology
{1110}
hide / edit[4] / print
ref: -0 tags: Seymour thesis electrode lithography fabrication date: 02-05-2012 17:35 gmt revision:4 [3] [2] [1] [0] [head]

Advanced polymer-based microfabricated neural probes using biologically driven designs.

  • References {1109}
  • Thermal noise from 280 um^2 or 170 um^2 gold recording sites much higher than PEDOT coated sites.
  • Used an interdigitated contact-free probe for measuring insulation impedance change. Very smart!
    • Water molecules will diffuse 15 um / minute in parylene (Yasuda, Yu et. al 2010).
    • In the frequency range critical for neural recording and stimulation, 500-5k, impedance moculus decline was small.
    • 1 hr soak at 60C.
  • Chapter 3 details 60-day soak of Parylene-C + reactive parylene insulation performance testing.
    • Regular parylene seems to work perfectly fine, no better than the PPX heat-treated devices.
    • Heat treatment does improve quality -- 200C in a vacuum oven for 2 days. (Li, Rodger et al 2005)
      • However -- this increases the brittleness.